PERFORMANCE
- Measurement Ranges: (parts per million) 0-10, 0-50, 0-100, 0-500, 0-1,000, 0-5,000, 0-10,000, and 0-20,000
- Accuracy¹: ±1% of full scale (± 5% of full scale on ranges ≤ 0-5ppm)
- ¹Stated at constant temperature and pressure
- Linearity: + 1% of full scale.
- Response Time: 90% of full scale response in less than 10 seconds (typical). The response time for ranges of 0-50 PPM or less depend to a great extent on the design of the sample delivery system including the materials used.
- Sensor Type: Long-life Ambient Temperature Electrochemical Sensor (Optional CO2 Resistant Sensor Available).
- Temperature Compensation: Standard.
- Operating Temperature Range: 40° to 104° F (5° to 40° C) <40° F (<5° C) use heated sensor enclosure >104° F (>40° C) cooling of sample gas/sensor required
- Warranty: Two years electronics, one year sensor
ELECTRICAL
- Series 3500 Loop Powered Trace Oxygen Transmitter: Input power 14-32 VDC
- Analog Output: 4-20 mADC with a loop resistance of 600 ohms @ 24 VDC (consult factory for other resistance values).
SAMPLE GAS CHARACTERISTICS
- Sample Flow Rate: 1.0 to 2.0 standard cubic feet per hour (SCFH). 0.5 to 1.0 liters/ minute (LPM).
- Sample Gas Temperature: 40° to 104° F (5° to 40° C)
- Sample Gas Pressure Limits: 0.1 to 1.5 psig (0.007 to 0.1 kg/cm2).
- Entrained Solids: <3 mg/ft³: no in-line filter required >3 mg/ft³: in-line filter is required
- Hydrocarbon Mist: <0.7 mg/ft³: no in-line filter required >0.7 mg/ft³: in-line filter is required
CONSTRUCTION
- Polycarbonate, rated NEMA 4X (IP66) without optional equipment. (Optional NEMA7 explosion-proof enclosure available)
- Dimensions: 5.5 in (139.9 mm) height. 8.8 in (223.5 mm) width. 3.4 in (86.4 mm) deep. Note: All dimensions are without optional equipment
- Gas Connections: 1/4” stainless steel compression fittings.